i-STAT Electronic Simulator Results Sheet PCT_iST.ESS v1. Title 04 Electronic Simulator QC Sheet Author: lnolan02 Created Date: 5/16/2012 4:13:10 PM
Purpose: i-STAT1 is tested with Electronic Quality Checks using an Electronic Simulator to confirm analyzer performance. Electronic Simulator Quality Control Regimen. Once every 24 hours when instrument not in use. Every 8 hours when instrument is in use.
Electronic Simulator. Running the external Electronic Simulator ensures the Internal Simulator cycle will not execute during the restoration cycle, which could lead to the premature
ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download
The schedule for the automatic internal Electronic Simulator can be set by the number of hours on the i-STAT Portable Clinical Analyzer and by the hours or tests on the i-STAT 1 Analyzer. Page 46: Operating Characteristics
Appendix 1: i-STAT Electronic Simulator Log i-STAT Serial #: Month: Year: Simulator Serial #: Site/Location: Date Time Result Pass/Fail **Only to be completed in event of simulator fail Operator **Failure Code or Letter **Action taken **Rpt Result Pass/Fail 1 P / F P / F 2 P / F P / F
Electronic Simulator i-STAT Printer Martel Printer See System Manual for instructions. For in vitro diagnostic use. 2 Intended Use: The i-STAT1 Analyzer is intended for use with i-STAT cartridges for the in vitro quantification of various analytes in whole blood. i-STAT 1 Downloader/Recharger i-STAT 1 Analyzer i-STAT Cartridges System Components 15
ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download | ManualsLib
View and Download Abbott I-STAT 1 system manual online. i-STAT 1 blood glucose meter pdf manual download.
Simulator attempt fails, contact APOC Technical Support. For additional information on running the electronic simulator, see Section 14 of the i-STAT 1 System Manual.
ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download
Page 208: Performing Electronic Simulator Test PERFORMING ELECTRONIC SIMULATOR TEST Procedure for Internal The internal Electronic Simulator test cycle is automatically activated when a Electronic Simulator cartridge is inserted after the customized interval is reached. If the analyzer passes the simulator test, the cartridge test cycle proceeds.
Quality control is performed on the analyser by an electronic simulator, which checks the non-disposable components of the i-STAT system. Testing of the analyser with liquid internal and external quality control samples provides a final full check of the total analyser system. 2. Specimen Collection